User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Tektronix 2232 100MHz Digital Storage Oscilloscope DATE: 12-Aug-93 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 69 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 HEAD Preliminary Procedure 1.002 DISP Connect the UUT IEEE cable to the Tektronix 2232 1.002 DISP Oscilloscope IEEE connector. 1.003 DISP Set the UUT VERTICAL controls as follows: 1.003 DISP 1.003 DISP ÿ MODE ........................................... CH1 1.003 DISP ÿ X-Y ............................... Off (button out) 1.003 DISP ÿ BW LIMIT ............................ On (button in) 1.003 DISP ÿ MODE ........................................... ALT 1.004 DISP Set the UUT CH1 VERTICAL controls as follows: 1.004 DISP 1.004 DISP ÿ POSITION .................................. Midrange 1.004 DISP ÿ VOLTS/DIV ...................................... 2mV 1.004 DISP ÿ VOLTS/DIV Variable ...................... CAL Detent 1.004 DISP ÿ AC-GND-DC ....................................... DC 1.005 DISP Set the UUT CH2 VERTICAL controls as follows: 1.005 DISP 1.005 DISP ÿ POSITION .................................. Midrange 1.005 DISP ÿ VOLTS/DIV ...................................... 2mV 1.005 DISP ÿ VOLTS/DIV Variable ...................... CAL Detent 1.005 DISP ÿ AC-GND-DC ....................................... DC 1.005 DISP ÿ INVERT ............................ Off (button out) 1.006 DISP Set the UUT HORIZONTAL controls as follows: 1.006 DISP 1.006 DISP ÿ POSITION .................................. Midrange 1.006 DISP ÿ MODE ............................................. A 1.006 DISP ÿ A & B SEC/DIV ................. (A & B LOCKED) 0.5ms 1.006 DISP ÿ SEC/DIV Variable ........................ CAL Detent 1.006 DISP ÿ B DELAY TIME POSITION .................... Fully CCW 1.006 DISP ÿ X10 Magnifier ........................ Off (knob in) 1.007 DISP Set the UUT A TRIGGER controls as follows: 1.007 DISP 1.007 DISP ÿ A TRIGGER MODE ............................ P-P AUTO 1.007 DISP ÿ SLOPE ............................... + (button out) 1.007 DISP ÿ LEVEL ..................................... Midrange 1.007 DISP ÿ A & B SOURCE ............................. VERT MODE 1.007 DISP ÿ A COUPLING .................................... NORM 1.007 DISP ÿ A EXT COUPLING .................................. AC 1.008 DISP Set the UUT B TRIGGER controls as follows: 1.008 DISP 1.008 DISP ÿ VAR HOLDOFF ................................... NORM 1.008 DISP ÿ SLOPE ............................... + (button out) 1.008 DISP ÿ LEVEL ..................................... Midrange 1.009 DISP Set the UUT STORAGE controls as follows: 1.009 DISP 1.009 DISP ÿ MODE ........................ NON-STORE (button out) 1.010 CALL Tektronix 2232 Sub1 1.011 CALL Tektronix 2232 Sub2 1.012 CALL Tektronix 2232 Sub3 1.013 CALL Tektronix 2232 Sub4 1.014 CALL Tektronix 2232 Sub5 1.015 CALL Tektronix 2232 Sub6 1.016 CALL Tektronix 2232 Sub7 1.017 CALL Tektronix 2232 Sub8 1.018 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.018 DISP {º This Completes the Verification º} 1.018 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.019 END